International Metal
Engineering |
|
|
|
|
|
|
|
|
|
|
MODEL 8080PA FIELD
MOUNTED 2 WIRE PRESSURE TRANSMITTER |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
n |
HIGH ACCURACY |
|
|
|
|
|
|
|
|
|
|
|
n |
4 :
1 RANGEABILITY |
|
|
|
|
|
|
|
|
|
|
n |
PRESSURE RANGE OF 0.025
BAR (10" H2O) |
|
|
|
|
|
TO 700 BAR (10,000 PSIG) |
|
|
|
|
|
|
|
|
|
|
n |
4 DIGIT LED DISPLAY,
ENGINEERING UNITS |
|
|
|
|
|
|
|
|
|
|
|
n |
CHOICE OF ALUMINUM OR
SS316 ENCLOSURE |
|
|
|
|
|
|
n |
EXPLOSION PROOF CERTIFIED |
|
|
|
|
|
|
|
|
n |
REMOTE MOUNT SENSOR FROM
ENCLOSURE |
|
|
|
|
|
|
n |
3 YEAR WARRANTY |
|
|
|
|
|
|
|
|
|
Introduction |
|
|
|
|
|
IME Model 8080PA is a
field mounted, 2 wire transmitter with a 4 digit |
|
|
LED
display and a 4 : 1 adjustability. The 8080PA is designed to cover |
|
|
a
wide range of level and pressure measurement and can be adjusted |
|
|
as
low as 0.025 Bar (10" H20) and up to a maximum of 700 Bar |
|
|
(10,000 PSIG). |
|
|
|
|
|
|
|
Description |
|
|
|
|
|
IME Model 8080PA
Transmitter is designed for use in process industries |
|
|
where vibration, inclement weather and
corrosive atmospheres prevail. |
|
|
The only material that is
exposed to the process liquid is SS316. The |
|
|
Electronics
are enclosed in a low copper epoxy coated Aluminum |
|
|
housing.
For more aggressive environments, a SS316 housing is |
|
|
optionally available. The
housings meet the requirements of NEMA4X, |
|
|
|
|
|
|
and are also certified
Explosion Proof by FM. The output current |
|
|
|
is linearised and can be
set to 4-20mA/ 20-4mA or any range within |
|
|
|
these
limit. The unit updates 3 times per second for display and |
|
|
|
4 times per second for
the output. The sensor can be remotely mounted |
|
|
|
up to 10 meters(30 feet)
from the Enclosure. |
|
|
|
|
|
|
|
Functional Specifications |
|
|
Turn On Time |
|
|
|
|
Process Fluid |
|
|
|
|
|
|
Performs within specification
approximately |
|
|
|
Liquid, Gas and Vapor |
|
|
|
|
2 seconds after power is applied to the
transmitter |
|
|
|
Output Signal |
|
|
|
Temperature Limits |
|
|
|
|
|
|
|
|
4-20 mA DC |
|
|
|
|
|
Process: -50°C to +125°C |
|
|
|
|
Power Supply |
|
|
|
Ambient: -50°C to +75°C |
|
|
|
|
|
|
|
|
|
|
|
12 to 45 Vdc |
|
|
|
|
|
Performance
Specifications |
|
|
|
|
Load Limitation |
|
|
|
Indication Accuracy |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
± 0.25% Full Scale, includes effects of
non- linearity, |
|
|
|
|
|
|
hysteresis and repeatability |
|
|
|
|
|
|
|
Stability |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
± 0.2% Full Scale Output for 12 months |
|
|
|
|
|
|
|
Temperature Effect |
|
|
|
|
According to: Rmax. (Ω) = (V
supply-12) / 0.02 |
|
|
|
|
|
|
|
Zero, Span Adjustment
& Configuration |
|
|
± 0.02% Full Scale Output per 20º C |
|
|
|
|
|
Electrical Connection |
|
|
Local Adjustment using Membrane
Switches |
|
|
|
|
|
|
Zero Adjustment Limits |
|
|
1/2"NPT, 3/4"NPT or M20 x 1.5 |
|
|
|
|
|
|
|
|
|
Process Connection |
|
|
|
|
|
|
|
|
|
|
|
1/2"NPT |
|
|
|
|
|
|
|
|
|
|
|
Electronic Housing |
|
|
|
|
|
|
|
|
|
|
IME MODEL 8080 Aluminum/ Stainless Steel
Instrument Enclosures |
|
|
|
|
|
Cover O-Rings |
|
|
|
|
|
|
|
|
Buna N |
|
|
|
|
|
|
|
|
|
Identification Plate |
|
|
|
|
|
|
|
|
|
|
Calibrated Span shall not be less then
0.25URL |
|
Stainless Steel SS304 |
|
|
|
|
|
and shall not exceed URL. |
|
|
Mounting |
|
|
|
|
|
Low Range Value shall not be below -URL |
|
|
|
|
|
|
|
Upper Range Value shall not be greater
than URL |
|
|
Wall, Panel or Pipe Mounting |
|
|
|
|
|
Indicator |
|
|
|
|
|
|
|
|
|
|
4 Digit LED indicator |
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
International Metal
Engineering Pte Ltd
http://www.intmet.com |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|